Surgeons-Don't Forget to Calibrate! Findings from a Sacral Nerve Test Stimulator

Etherson, K.; Davidson, A.; Yiannakou, Y.; Mason, J.

Surgeons-Don't Forget to Calibrate! Findings from a Sacral Nerve Test Stimulator Thumbnail


Authors

K. Etherson

A. Davidson

Y. Yiannakou

J. Mason



Abstract

Introduction: Sacral nerve stimulation testing (TSNS) for chronic constipation is not accurately predictive of a long-term response1. The decision to implant a permanent device relies on these results. Aims/Background: We recognised that the testing stimulator was an analogue device with potentially inaccurate dial settings. We sought verification of the output waveform. Method: 19 test stimulators were connected to a cross-calibrated oscilloscope. The output Frequency (f), and Pulse Width (pw) of the waveforms generated were measured according to: run 1) the physician's best attempt to set the dials correctly (pw=210 µ Sec, f=14 Hz), and run 2) the closest dial increment to these settings (pw=200 µ Sec, f=10 Hz). Output Voltage (V) was measured in run 3 at dial increments of 0V, 1V, 2V, 5V, and 10V. Results: We assumed an acceptable margin of error of 20% in runs 1 and 2, and 0.5V in run 3. There was a marked range of frequency values; run 1) 10.6 to 29.0 Hz (26% failed), and run 2) 7.9 to 13.0 Hz (11% failed). Findings for pulse width were similarly variable; run 1) 242 to 326 µ Sec (89% failed), and run 2)215 to 274 µ Sec (63% failed). All devices had a residual positive voltage at zero(range:0.29 to 1.00V), and the failure rates at 0,1,2,5 and 10V were 53%, 100%, 100%, 68% and 47% respectively. Conclusion: All fields of clinical practice and research have their instrumentation which requires calibration to provide verifiable readings. Failure to calibrate during TSNS results in patients receiving variable stimulation, potentially reducing the clarity of research findings, and may be a factor in the poor predictive power of testing in chronic constipation.

Citation

Etherson, K., Davidson, A., Yiannakou, Y., & Mason, J. (2013). Surgeons-Don't Forget to Calibrate! Findings from a Sacral Nerve Test Stimulator. Gut, 62(Suppl 2), A27-A28. https://doi.org/10.1136/gutjnl-2013-305143.64

Journal Article Type Article
Publication Date Aug 1, 2013
Deposit Date Sep 4, 2013
Publicly Available Date Jan 13, 2014
Journal Gut
Print ISSN 0017-5749
Electronic ISSN 1468-3288
Publisher BMJ Publishing Group
Peer Reviewed Peer Reviewed
Volume 62
Issue Suppl 2
Pages A27-A28
DOI https://doi.org/10.1136/gutjnl-2013-305143.64

Files

Accepted Journal Article (187 Kb)
PDF

Copyright Statement
This article has been accepted for publication in Gut. The definitive copyedited, typeset version Etherson, K. and Davidson, A. and Yiannakou, Y. and Mason, J. (2013) 'Surgeons-don't forget to calibrate ! findings from a sacral nerve test stimulator.', Gut., 62 (Suppl 2). A27-A28 is available online at: http://dx.doi.org/10.1136/gutjnl-2013-305143.64




You might also like



Downloadable Citations