Feasibility of Macular Integrity Assessment (MAIA) microperimetry in children: Sensitivity, reliability, and fixation stability in healthy observers

Jones, P.R.; Yasoubi, N.; Nardini, M.; Rubin, G.S.

Feasibility of Macular Integrity Assessment (MAIA) microperimetry in children: Sensitivity, reliability, and fixation stability in healthy observers Thumbnail


Authors

P.R. Jones

N. Yasoubi

G.S. Rubin



Abstract

Purpose: To assess the feasibility of macular integrity assessment (MAIA) microperimetry (MP) in children. Also to establish representative outcome measures (differential light sensitivity, fixation stability, test–retest reliability) for children without visual impairment. Methods: Thirty-three adults and 33 children (9–12 years) were asked to perform three monocular MAIA examinations within a single session (dominant eye only). Results: Children exhibited poorer test–retest reliability than adults for measures of both mean sensitivity (95% coefficient of repeatability [CoR95] = 2.7 vs. 2.3 dB, P = 0.036) and pointwise sensitivity (CoR95 = 6.2 vs. 5.7 dB, P < 0.001). Mean sensitivity was lower in children (27.6 vs. 29.5 dB, P < 0.001), and fixation stability was poorer (95% bivariate contour ellipse area [BCEA95] = 4.58 vs. 1.14, P < 0.001). Mean sensitivity was negatively correlated with fixation stability (r = −0.44, P < 0.001). Both children and adults exhibited substantial practice effects, with mean sensitivity improving by 0.5 dB (adults) and 0.9 dB (children) between examinations 1 and 2 (P ≤ 0.017). There were no significant differences between examinations 2 and 3 (P ≥ 0.374). Conclusions: Microperimetry is feasible in 9- to 12-year-old children. However, systematically lower sensitivities mean that the classification boundary for “healthy” performance should be lowered in children, pending development of techniques to improve attentiveness/fixation that may reduce or remove this difference. High measurement variability suggests that the results of multiple tests should be averaged when possible. Learning effects are a potential confound, and it is recommended that the results of the first examination be discarded.

Citation

Jones, P., Yasoubi, N., Nardini, M., & Rubin, G. (2016). Feasibility of Macular Integrity Assessment (MAIA) microperimetry in children: Sensitivity, reliability, and fixation stability in healthy observers. Investigative Ophthalmology & Visual Science, 57(14), 6349-6359. https://doi.org/10.1167/iovs.16-20037

Journal Article Type Article
Acceptance Date Sep 9, 2016
Online Publication Date Nov 29, 2016
Publication Date Nov 29, 2016
Deposit Date Sep 9, 2016
Publicly Available Date Dec 8, 2016
Journal Investigative Ophthalmology & Visual Science
Print ISSN 0146-0404
Electronic ISSN 1552-5783
Publisher Association for Research in Vision and Ophthalmology
Peer Reviewed Peer Reviewed
Volume 57
Issue 14
Pages 6349-6359
DOI https://doi.org/10.1167/iovs.16-20037