We present results from cryogenic tests of volume-phase holographic (VPH) gratings at ∼100 K. The aims of these tests are to see whether the diffraction efficiency as a function of wavelength is significantly different at a low temperature from that at room temperature and to see how the performance of a VPH grating is affected by a number of thermal cycles. We have completed ten cycles between room temperature and 100 K and find no clear evidence that the diffraction efficiency changes with temperature or with a successive thermal cycle.

Citation

Tamura, N., Murray, G., Luke, P., Blackburn, C., Robertson, D., Dipper, N., …Allington-Smith, J. (2006). Cryogenic tests of volume-phase holographic gratings: results at 100K. Applied optics, 45(23), 5923-5928. https://doi.org/10.1364/ao.45.005923

Journal Article Type Article
Publication Date 2006-08
Deposit Date Feb 21, 2008
Publicly Available Date Nov 10, 2010
Journal Applied Optics
Print ISSN 0003-6935
Publisher The Optical Society
Peer Reviewed Peer Reviewed
Volume 45
Issue 23
Pages 5923-5928
DOI https://doi.org/10.1364/ao.45.005923
Keywords Volume holographic gratings, Spectroscopy, Infrared.

Files

Published Journal Article (317 Kb)
PDF

Copyright Statement
© 2006 Optical Society of America.




Downloadable Citations